Subject:

Re: Weird rings in Relion 2D classificaction

From:

Sjors Scheres <[log in to unmask]>

Reply-To:

Sjors Scheres <[log in to unmask]>

Date:

Mon, 27 Oct 2014 22:13:17 -0000

Content-Type:

text/plain

Parts/Attachments:

Parts/Attachments

text/plain (199 lines)

Yes, 0.8 would actually be very conservative, so I was somewhat surprised.
Useful threshold values are often around 0.2-0.7. 0.08 indeed sounds very
low.
S


> Sorry there was a typo - 0.08 was what I used, not 0.8. That should make
> more sense!
>
> On Mon, Oct 27, 2014 at 9:05 AM, Liz Montabana <[log in to unmask]>
> wrote:
>
>> Hi Sjors
>>
>> Thanks for the advice. I had originally gone through the described
>> FOM/particle threshold picking with a few micrographs and ended up
>> picking
>> a threshold of 0.8, which looked like it was getting mostly what I would
>> have picked out as particles by eye. I guess I would be a pretty greedy
>> manual picker too! I'll go back and pick something more conservative and
>> see how that turns out. For the auto-picking I was low-pass filtering to
>> 20A, but I'll change that to 30 for this round and see what happens.
>>
>> Do you think that it would be better to not mask your original mini-set
>> of
>> manual picked particles when doing 2D class averages, so when using them
>> for templates there is no chance of that affecting the auto-picking?
>>
>> thanks
>> Liz
>>
>>
>>
>>
>> On Sat, Oct 25, 2014 at 4:52 AM, Sjors Scheres
>> <[log in to unmask]>
>> wrote:
>>
>>> Dear Liz,
>>>
>>> Don't apologize for this excellent and well-documented question!
>>>
>>> The rings you observe are a typical sign of Einstein-from-noise
>>> artifacts
>>> that can occur with (any) template-based particle picking. I attach a
>>> PDF
>>> of the manuscript I wrote on the auto-picking procedure in RELION. Note
>>> this hasn't been back from peer-review yet, so for the moment just
>>> consider it as my own opinion! :-) We can observe the same effects with
>>> many more data sets. The cause is a too low threshold for the
>>> auto-picking
>>> algorithm, which then reproduces the templates from noise areas in the
>>> micrographs. I suppose you used the smaller diameter to calculate the
>>> 2D
>>> classes that you used as templates for auto-picking? The normalization
>>> in
>>> relion (and many other packages) makes that the average background is
>>> often slightly negative (i.e. black); this is caused by the presence of
>>> particles (with white signal) in the background area. This results in
>>> 2D
>>> class averages that have a white ribosome and a slightly negative
>>> (black)
>>> background. When you use those for auto-picking and you are too
>>> "greedy",
>>> i.e. you set the threshold too low, then you start up picking noise
>>> that
>>> also describes your templates. 2D class averaging with those large,
>>> noisy
>>> data sets then gives rise to the typical black rings in the class
>>> averages
>>> (see also Fig 5 in the attached paper). The good class averages (e.g.
>>> first row, number 5 and second row number 1 &2 in your figure) do not
>>> show
>>> these template-induced artefacts.
>>> You can now do 2 things: 1) hope 2D classification deals with this, and
>>> just throw away all Einstein-from-noise classes; or 2) (probably
>>> better):
>>> re-run auto-picking with a higher threshold and then re-run 2D
>>> classification.
>>> Many of your good classes seem to represent a similar view, so you may
>>> have a strong preferred orientation, but this is better assessed once
>>> you've cleaned up your data from the false positives of the
>>> auto-picking
>>> approach.
>>>
>>> The Einstein-from-noise artifacts are an important reason why you
>>> should
>>> ALWAYS use low-pass filtered templates for auto-picking. Probably using
>>> beyond 30A information is not useful anyway, and information that did
>>> not
>>> go into the process can also not come out as an Einstein-from-noise
>>> artifact. Using high-resolution templates for auto-picking is probably
>>> what went wrong in the HIV trimer structure that caused such a stir-up
>>> last year. Both Richard Henderson and Marin van Heel wrote excellent
>>> commentaries in PNAS on that issue. See
>>> http://www.ncbi.nlm.nih.gov/pubmed/24106306 and
>>> http://www.ncbi.nlm.nih.gov/pubmed/24106301.
>>>
>>> HTH,
>>> Sjors
>>>
>>>
>>> > Hi All
>>> >
>>> > I hope this is not too bone-headed of a question. I am new to using
>>> Relion
>>> > (1.3), and I am doing 2D classification and I have been getting some
>>> > results that look suspicious. I am using my own DD data set of yeast
>>> 80S
>>> > particles. I followed the steps outlined in "Tutorial (The quickest
>>> way
>>> to
>>> > learning RELION-1.3) ", for ctf-correction, autopicking, etc. - Now I
>>> am
>>> > trying to do some particle pruning with the 2D classification. There
>>> are
>>> > two things going on:
>>> > 1) There are two highly populated classes (40% of ~60k particles)
>>> that
>>> > look the same - they seem to be neither junk nor good classes.
>>> >
>>> > 2) There are two visible rings for each class - one is the soft mask,
>>> as I
>>> > would expect, but the other seems to be related to the rotation of
>>> the
>>> > particle during alignment - I've attached an image - it seems to
>>> exist
>>> on
>>> > both good and bad classes - these are sorted by class distribution. I
>>> > can't figure out what causes this, but I feel like its a good
>>> indicator
>>> > I've picked some bad parameter. When I did 2D Classification on a
>>> very
>>> > small manual picked set with a very slightly smaller mask (350 A vs
>>> 360
>>> A)
>>> > I did not see the same thing. My box size is 300 pix = 396 Ang.
>>> >
>>> > Thanks for any help -  please excuse such a possible inane question!
>>> > Liz Montabana
>>> >
>>> > Here is the command:
>>> > mpirun -np 240 --bynode `which relion_refine_mpi` --o
>>> > Class2D/autopick_sort2 --i particles_autopick_rd2.star
>>> --particle_diameter
>>> > 360 --angpix 1.32 --ctf  --iter 25 --tau2_fudge 2 --K 300
>>> > --flatten_solvent  --oversampling 1 --psi_step 10 --offset_
>>> > range 5 --offset_step 2 --norm --scale  --j 1 --memory_per_thread 8
>>> >
>>> > Class2D/autopick_sort2.out
>>> >
>>> > and settings:
>>> > Class2D/autopick_sort2.gui_class2d.settings
>>> > is_continue == false
>>> > Output rootname: == Class2D/autopick_sort2
>>> > Continue from here:  ==
>>> > Input images STAR file: == particles_autopick_rd2.star
>>> > Number of classes: == 300
>>> > Do CTF-correction? == Yes
>>> > Have data been phase-flipped? == No
>>> > Ignore CTFs until first peak? == No
>>> > Number of iterations: == 25
>>> > Regularisation parameter T: == 2
>>> > Mask individual particles with zeros? == No
>>> > Limit resolution E-step to (A):  == -1
>>> > Perform image alignment? == Yes
>>> > In-plane angular sampling: == 5
>>> > Offset search range (pix): == 5
>>> > Offset search step (pix): == 1
>>> > Number of MPI procs: == 240
>>> > Number of threads: == 1
>>> > Available RAM (in Gb) per thread: == 8
>>> > Submit to queue? == Yes
>>> >
>>> >
>>> >
>>>
>>>
>>> --
>>> Sjors Scheres
>>> MRC Laboratory of Molecular Biology
>>> Francis Crick Avenue, Cambridge Biomedical Campus
>>> Cambridge CB2 0QH, U.K.
>>> tel: +44 (0)1223 267061
>>> http://www2.mrc-lmb.cam.ac.uk/groups/scheres
>>
>>
>>
>


-- 
Sjors Scheres
MRC Laboratory of Molecular Biology
Francis Crick Avenue, Cambridge Biomedical Campus
Cambridge CB2 0QH, U.K.
tel: +44 (0)1223 267061
http://www2.mrc-lmb.cam.ac.uk/groups/scheres

Top of Message | Previous Page | Permalink

JiscMail Tools


RSS Feeds and Sharing


Advanced Options


Archives

April 2024
March 2024
February 2024
January 2024
December 2023
November 2023
October 2023
September 2023
August 2023
July 2023
June 2023
May 2023
April 2023
March 2023
February 2023
January 2023
December 2022
November 2022
October 2022
September 2022
August 2022
July 2022
June 2022
May 2022
April 2022
March 2022
February 2022
January 2022
December 2021
November 2021
October 2021
September 2021
August 2021
July 2021
June 2021
May 2021
April 2021
March 2021
February 2021
January 2021
December 2020
November 2020
October 2020
September 2020
August 2020
July 2020
June 2020
May 2020
April 2020
March 2020
February 2020
January 2020
December 2019
November 2019
October 2019
September 2019
August 2019
July 2019
June 2019
May 2019
April 2019
March 2019
February 2019
January 2019
December 2018
November 2018
October 2018
September 2018
August 2018
July 2018
June 2018
May 2018
April 2018
March 2018
February 2018
January 2018
December 2017
November 2017
October 2017
September 2017
August 2017
July 2017
June 2017
May 2017
April 2017
March 2017
February 2017
January 2017
December 2016
November 2016
October 2016
September 2016
August 2016
July 2016
June 2016
May 2016
April 2016
March 2016
February 2016
January 2016
December 2015
November 2015
October 2015
September 2015
August 2015
July 2015
June 2015
May 2015
April 2015
March 2015
February 2015
January 2015
December 2014
November 2014
October 2014
September 2014
August 2014
July 2014
June 2014
May 2014
April 2014
March 2014
February 2014
January 2014
December 2013
November 2013
October 2013
September 2013
August 2013
July 2013
June 2013
May 2013
April 2013
March 2013
February 2013


JiscMail is a Jisc service.

View our service policies at https://www.jiscmail.ac.uk/policyandsecurity/ and Jisc's privacy policy at https://www.jisc.ac.uk/website/privacy-notice

For help and support help@jisc.ac.uk

Secured by F-Secure Anti-Virus CataList Email List Search Powered by the LISTSERV Email List Manager