We have noted a severe problem with kit lot 232 Free T3 giving internal QC results that are in spec but VERY low patient results (eg below 1.5 pmol/L). Problem corrected with another Lot no. Anyone else using this Lot? Noted same or similar problem? Problem "being referred" to the States by DPC but not sure if other customers aware or problem affects all of that Lot.