Hello everyone,
Thanks for the suggestion Claire: I think I did (try to) send HTML, but the
tables which I sent earlier did not make it through our email server in a
respectable format for whatever reason.
I will soon send out some cleared versions of the Excel files to those who
have requested them.
The info (in the case of an eppendorf dispenser)is as (very roughly)
presented below: I have used blue for column heading, red for calculated
results, and green for items entered by the operator. In the real thing, all
the data for a particular check is on a single line.
Hopefully this will arrive looking slightly better than the looooong string:
if not...?
Richard Scarff
Serial Date Operator Temp Tip: Setting
oC "1"=vol (1 to 5)
1 29/06/1999 rs 22.3 1000 5
Corrected Average SD (uL) CV% Pipette Accuracy Precision
volume (uL) Error% (Epp specs) (Epp
specs)
5001.37 2.20 0.044 0.027 ok ok
Biochem specs: Data Weighings ... (max 50)
Acc/Prec Error
a p ok 4.9895 4.9893 4.9910 4.9925 4.9874 4.9878 4.9925
4.9934 4.9877 4.9891
-----Original Message-----
From: Taylor, Claire
Sent: Friday, 14 November 2003 10:59
To: Scarff, Richard
Subject: RE: Pipette checks
------ACB discussion List Information--------
This is an open discussion list for the academic and clinical
community working in clinical biochemistry.
Please note, archived messages are public and can be viewed
via the internet. Views expressed are those of the individual and
they are responsible for all message content.
ACB Web Site
http://www.acb.org.uk
List Archives
http://www.jiscmail.ac.uk/lists/ACB-CLIN-CHEM-GEN.html
List Instructions (How to leave etc.)
http://www.jiscmail.ac.uk/
|