> In danish and nordic EQAS-programmes, our Dimension Rxl analysers
> consistently give 10-12 % too low Magnesium results in the range 0.6 -
> 1.5 mmol/l. Dade-Behring has not devised a solution to the problem in
> spite of many complaints. Do UK-users of the machines realise the same
> problem and what do you do about it?
UK NEQAS performance for 8 Dimension instruments shows a
negative bias relative to their all-laboratory trimmed mean (ALTM)
target value, currently averaging -7% (range -3% to -9%), and in
most cases the bias appears to be greater at concentrations below
1.5 mmol/L.
This was previously drawn to the attention of DADE Behring, when
the problem appeared to be worse and placed several users
outside the limits of acceptable performance, but with no definitive
outcome
David Bullock
Dr David Bullock
Director, Wolfson EQA Laboratory
P O Box 3909, Birmingham B15 2UE, U K
FAX: 0121 414 1179 [+44 121 414 1179]