Hi,
I noticed recently that maxfilter 2.0 origin fit to isotrak has a flaw:
if landmark points are digitized twice (also as extra points), the
program skips the fit and sets the origin erroneosuly to 0,0,0.
Anyway, if the program reports 0,0,0 and you are not processing
phantom data, the remedy is to set origin manually.
I am finalizing maxfilter version 2.1, where this pronlem (and numerous
others) are fixed. We are aiming to release it at the end of August.
Sorry for the inconvenience, and best regards
Jukka nenonen
================================
Dr. Jukka Nenonen
Manager, Method development
Elekta Neuromag Oy
Street address: Siltasaarenkatu 18-20A, Helsinki, Finland
Mailing address: P.O. Box 68, FIN-00511 HELSINKI, Finland
Tel: +358 9 756 240 85 (office), +358 400 249 557 (mobile),
+358 9 756 240 11 (fax), +358 9 756 2400 (operator)
E-mail: [log in to unmask]
http://www.neuromag.com
|