Hi Max, the Siemens fieldmap sequence stores the magnitude data in one folder and the phase image in the other (Phase and Magnitude Data is always stored in two different folders for the Standard Siemens Sequences). So what you need to to is convert both folders, load one of the magnitude images and the phase image into the fieldmap toolbox, enter the two echo times at which your fieldmap was measured and it should work. Hope this helps, Benjamin Maximilien Chaumon schrieb: > Hi SPMers > > I'm trying to realign & unwarp EPI images because I'm interested on > orbitofrontal activity and would like to be sure I map correctly this part. > I'd thus like to use the field maps I've been acquiring just before my > functionals with the same parmeters. The thing is that I only have a single > nii file in my 'fieldmap' directory after dicom import (The scanner I'm > using is a Siemens Trio 3T). This file does not give me phase & amplitude or > real & imaginary that I could feed the fieldmap toolbox with in order to > obtain a vdm file to feed my realign & unwarp process. I noticed that I get > actually two fieldmap directories per sequence I ran. I guess they could be > phase & amplitude or real & imaginary but I don't know which is which... > > So my question is simple: how do I get two measures out of my sequences > (either phase & amplitude or real & imaginary) or if they are these two > separate directories how do I know which is which? should I do a different > dicom import? > > Or am I completely wrong, trying to do something that doesn't make sense? > > Thanks! > Max > > -- Benjamin Bender Abteilung Diagnostische und Interventionelle Neuroradiologie Universitätsklinikum Tübingen Hoppe-Seyler- Str. 3 72076 Tübingen Tel. (07071) 29-85397