Hi Nina,
If you want to compare the intensities of spectra which were exported
from Topspin into another format, then make sure these spectra were
all processed with the same value NR fo the nc_proc parameter, using
the command "xfb nc_proc NR".
Hope that helps,
Jakob
On Fri, Nov 25, 2011 at 15:16, Nina Luckgei <[log in to unmask]> wrote:
> Hi all,
>
> I wanted to compare signal intensities of two different spectra. However their NC_proc parameter are not equal. I just found an old conversation (02/2008) in the mailing list about this topic which says that this parameter might be incorporated into a new (back then) version. Were you able to integrate the NC_proc parameter so that the peakvolume (or signal intensity) is not dependent on this parameter? If yes, is it the parameter 'scale' in the spectra menu?
>
> Thank you,
> Nina
>
|