> >Usually the fix is as simple as swapping the 1H Ref Exp Dims.
>
> OK. In my 3dNnoesy I have the next:
> ExpDim: 1 (Hn) 2 (N) 3 (Hx)
> RefExpDim: 1 (1H) 2 (15N) 3 (1H)
> RefMeasurements: 1,3 2 4
>
> But I guess that RefMeasurements should be 2,4 3 1 and there is no way to
> set it.
That would be sensible, but for data model reasons H_H[N] is stored in the
same reference prototype as H[N]_H, only 'reversed'. I now realise that
this means that the Ref Measurement numbers appear backwards to what you'd
expect. - I will fix Analysis to put this right. However, I reckon you
should try Ref Measurements 4 2 1,3, i.e. swap the proton dims as I
suggest.