This sounds ok. SIENA estimated shift to subvoxel accuracy using an
extended profile perpendicular to the edge. When further regularised
via a) smoothing and b) cross-subject averaging then yes this sounds
reasonable.
Cheers.
On 12 Sep 2007, at 16:28, Matthew Kempton wrote:
> Dear SIENA users,
>
> Thanks for the information Steve, I have now obtained the correct flow
> values.
> However I have noticed that they are still quite small.
>
> From our experiment we expect only a small change in boundary
> shifts. For
> seven subjects we had a peak mean flow value of 0.1mm (SD=0.01 mm)
> in the
> lateral ventricle. Although this is highly significant, is SIENA
> really
> able to be sensitive to such a small change (original MRI voxel
> size was
> 0.94 x 0.94 x 1.5mm).
>
>
> Thanks,
>
> Matthew
------------------------------------------------------------------------
---