Definitely sounds like a bug! Can you email me the project files (the
ABC.xml and ABC/*/*.xml files)? Not the spectra data files, I shouldn't
need those to look at this. And tell me how to reproduce it on that
project.
Thanks, Wayne
On Mon, 21 Aug 2006, Christoph Brockmann wrote:
> Wayne Boucher wrote:
> > Hmmm, sounds like a bug. What happens (as of version 1.0.11) in
> > orthogonal dimensions is that an extra half point of thickness is used
> > both for displaying and selecting peaks. At least that's the theory.
> > (And this is in addition to whatever thickness you might have added in the
> > peak "find params" dialog.) This extra half a point was added because of
> > the problem that interpolation to find the peak maximum can cause the
> > position to be up to half a point from the grid point, and this could mean
> > that peaks being picked might not be shown. But if you are able to not
> > see something but can pick it then that sounds like a bug. How many
> > points in the z direction in the HSQC-NOESY is 0.2 ppm equal to?
> 1 point is approximately 0.022ppm, so its about 9 points. the "extra thickness" in the peak find parameters is set to 0.
>
> thanks,
> Christoph
>
> ---------------------------------------------------------
> Dr. Christoph Brockmann
> Division of Structural Studies
> MRC Laboratory of Molecular Biology
> Hills Road
> Cambridge, CB2 2QH
> phone: +44-1223-40-2261 or +44-1223-40-2068
> e-mail: [log in to unmask]
> ---------------------------------------------------------
>
|