This is my current record of Cs values on TFS systems for bio, maybe TFS people on this list can confirm these:

Microscope pole type Cs [mm]

Tecnai 10 High-Contrast 4.3
Tecnai 12   Bio-Twin 6.3

Tecnai 12 Twin 2.26
Tecnai (F)20 Twin 2.26
Tecnai (F)30/Polara Twin 2.26
Glacios C-twin 2.7
Tecnai Arctica C-twin 2.7
Talos Arctica C-twin 2.7
Titan Krios & Halo C-twin 2.7

Best,

Wim Hagen
EMBL Heidelberg


On 13. Jun 2019, at 13:59, Grassucci, Robert <[log in to unmask]> wrote:

Our Polara has a Cs of 2.26 mm.  This was actually measured before it was installed but most twin lenses are said by Philips/FEI/Thermofisher  to have a Cs of 2.0 mm.
Bob
 
From: Collaborative Computational Project in Electron cryo-Microscopy [mailto:[log in to unmask]] On Behalf Of sunyeping
Sent: Wednesday, June 12, 2019 4:07 AM
To: [log in to unmask]
Subject: [ccpem] what is the spherical aberration of TF30 Polara electron microscope (FEI)
 
Dear everyone, 
 
I am trying to process microscophs data collected from TF30 Polara electron microscope with relion. The data is downloaded from EMPIAR. In the "CTF estimation" work type, the I/O tag interface require to input the  spherical aberration. Do you know what is the spherical aberration value of FEI TF30 Polara electron microscope?
 
Best regards
 
Yeping
 