Hi Gareth,
This comes down to the grade of glassware (class A or B). When I looked into this last year, Class A glassware complied with the UKAS accreditation and traceability standards. As such, grade B lab glassware could be calibrated in-house against grade (class) A glassware with documented records of the calibration data (this requires you to buy in at least one set of grade A glassware if you don’t already have any). Unless this has changed since a year ago (which it seems to do from lab-to-lab), this satisfied UKAS.
Regards,
Gary
Dr Gary Woodward PhD
Principal Clinical Scientist
SAS Steroid Endocrinology
Universtry College Hospitals
60 Whitfield Street London W1T 4EU
---------------------------------------------------------------------------------------------------------
Tel +44 (0) 20 344 72965
From: Clinical biochemistry discussion list [mailto:[log in to unmask]] On Behalf Of Gareth Davies (BCUHB - Pathology)
Sent: 01 November 2016 11:41
To: [log in to unmask]
Subject: Pipette Question
Good morning,
Are any ISO accredited labs using volumetric glassware, specifically pipettes for making up calibrators, IQC material etc?
Our automatic pipettes are regularly checked by an ISO-accredited company but we don’t routinely have our glass pipettes externally checked. Has this ever been brought up during an ISO inspection and if so, what were the solutions?
Many thanks.
Gareth
Gareth Davies
Biocemegydd Clinigol/Principal Clinical Biochemist
Bwrdd Iechyd Prifysgol Betsi Cadwaladr University Health Board
Ysbyty Maelor Wrecsam/Wrexham Maelor Hospital
Wrecsam/Wrexham LL13 7TD
01978 725345
P Please consider the environment before printing this email or its attachment(s)
Cymraeg
Rhybudd Ebost (2010) - Bwrdd Iechyd Prifysgol Betsi Cadwaladr
Fe'ch cynghorir i ddarllen rhybydd ebost Bwrdd Iechyd Prifysgol Betsi Cadwaladr (a'i argraffu er mwyn cyfeirio ato yn y dyfodol). Gellir dod o hyd iddo yn y lleoliad canlynol
http://www.wales.nhs.uk/sitesplus/861/tudalen/47230
English
Betsi Cadwaladr University Health Board - Email Notice (2010)
You are advised to read (and print for future reference) the Betsi Cadwaladr University Health Board e-mail notice which can be found at this location
http://www.wales.nhs.uk/sitesplus/861/page/47229
Betsi Cadwaladr University Health Board is the operational name of Betsi Cadwaladr University Local Health Board
------ACB discussion List Information-------- This is an open discussion list for the academic and clinical community working in clinical biochemistry. Please note, archived messages are public and can be viewed via the internet. Views expressed are those of the individual and they are responsible for all message content. ACB Web Site http://www.acb.org.uk Green Laboratories Work http://www.laboratorymedicine.nhs.uk List Archives http://www.jiscmail.ac.uk/lists/ACB-CLIN-CHEM-GEN.html List Instructions (How to leave etc.) http://www.jiscmail.ac.uk/
------------------------------------------------------------------------ This email is confidential and is intended solely for the person or entity to whom it is addressed. If this is not you, please forward the message to [log in to unmask] We have scanned this email before sending it, but cannot guarantee that malicious software is absent and we shall carry no liability in this regard. We advise that information intended to be kept confidential should not be sent by email. We also advise that health concerns should be discussed with a medical professional in person or by telephone. NHS 111 can also provide advice. We shall not be liable for any failure to follow this advice. University College London Hospitals NHS Foundation Trust (UCLH).