Suggest consider notifying MHRA.
Ian
From: Clinical
biochemistry discussion list [mailto:[log in to unmask]] On Behalf Of
Mackenzie, Paul
Sent: 14 July 2015 09:50
To: [log in to unmask]
Subject: FW: Anomalous results using POCT Chirus HCG meter - lot 615A161
Dear Mailbase
Warning to Chirus Biosign urine HCG users.
We have experienced anomalous pregnancy test results with BioSign RC hCG POCT Urine testing cassettes
lot 615A161 when read on the DXpress reader -
i.e. obtained negative results which have confirmed positive by serum BHGC on mainline analyser.
This is a lot specific issue.
Tony Blackman BSc MSc
Point of Care Testing
Principal Biomedical Scientist
Pathology First
01268 524900 ext. 8382
===============================================================
This email has been processed by the Southend Hospital mail gateway. Any executable files or known viruses have been removed - We do not accept or send emails larger than 10MB. Confidential information in this email must be handled in accordance with the Data
Protection Act 1998. Details of all emails are archived for seven days. If this message has been sent to you in error please discard the email, notify the sender and do not open any attachments or act on the contents. For more information contact the IT Department,
Southend University Hospital NHS Foundation Trust on +44 1702 385 678
------ACB discussion List Information-------- This is an open discussion list for the academic and clinical community working in clinical biochemistry.
Please note, archived messages are public and can be viewed via the internet. Views expressed are those of the individual and they are responsible for all message content. ACB Web Site
http://www.acb.org.uk Green Laboratories Work
http://www.laboratorymedicine.nhs.uk List Archives
http://www.jiscmail.ac.uk/lists/ACB-CLIN-CHEM-GEN.html List Instructions (How to leave etc.)
http://www.jiscmail.ac.uk/