Hi Kambiz,

This is an unfortunate bug in maxfilter version 2.2.15. It occurs when:

-        There is discrepancy of isotrak and initial hpifit in 1-2 coils, and

-        HPI amplitude extraction or coil fitting fails repeatedly within the tSSS buffer (here 4 s).

We have a fix coming later this year, but I don’t know the exact schedule yet.

You can try: 1) option hpicons to improve the initial consistency, 2) skip the problematic time interval(s), or 3) use previous version (2.2.10).

Installation of the new release wipes out 2.2.10, but I can send the binary by email and you can execute it e.g. from $(HOME)/bin.

Please report these problems also to [log in to unmask]<mailto:[log in to unmask]>.

Best regards, Jukka




Jukka Nenonen | Method Development Manager
Elekta Oy
Box 34, FI-00531 Helsinki, Finland
Visiting Address: Siltasaarenkatu 18-20 A, Helsinki
Office: +358 9 756 24085 | Mobile: +358 400 249 557 | Fax: +358 9 756 24011
[log in to unmask]<mailto:[log in to unmask]> | www.elekta.com<http://www.elekta.com>

[cid:image348393.JPG@87b261d1.44b8e135]
        [cid:image09bb05.PNG@ca124e62.47ae626a]


From: Announcement for the Neuro MEG list [mailto:[log in to unmask]] On Behalf Of Kambiz Tavabi
Sent: 1. huhtikuuta 2014 22:44
To: [log in to unmask]
Subject: Maxfilter exits with code 11 - unable to extract HPI amp, determine transformation

Hi all,
I am using MF version 2.2.15 on a MEG(306)+EEG(60) channels data set acquired at 1 kHz, with 0.1-300 Hz band filter.
I am encountering a Maxfilter failure (exit code 11) roughly halfway through a combined tSSS and movement compensation procedure. Before exit the program spits out:

Note! Discrepancy of HPI coil 3 isotrak and hpifit is 10.4 mm!
Note! Discrepancy of HPI coil 4 isotrak and hpifit is 6.8 mm!
Warning: HPI amp extraction failed (t=325.000)!
Warning: 2/4 acceptable hpi fits found, cannot determine the transformation! (t = 349.000)
Warning: 1/4 acceptable hpi fits found, cannot determine the transformation! (t = 350.000)
Warning: 1/4 acceptable hpi fits found, cannot determine the transformation! (t = 351.000)
Warning: 0/4 acceptable hpi fits found, cannot determine the transformation! (t = 352.000)
Simplex: nr of eval (1001) exceeds maximum (1000)!
Warning: 0/4 acceptable hpi fits found, cannot determine the transformation! (t = 353.000)
Simplex: nr of eval (1001) exceeds maximum (1000)!
Warning: 0/4 acceptable hpi fits found, cannot determine the transformation! (t = 354.000)
Simplex: nr of eval (1002) exceeds maximum (1000)!
Warning: 0/4 acceptable hpi fits found, cannot determine the transformation! (t = 355.000)
Program died due to signal 11, all output files were deleted!
Core dump disabled.

The Maxfilter command and relevant arguments are: maxfilter-2.2 -v -f ... -o ... -st 4 -corr 0.98 -in 8 -out 3 -regularize in -hpie 20 -hpig 0.8 -origin 0 0 40  -movecomp -hpisubt amp -hpistep 200 -frame head. I'v also tried using the arguments: -hpicoils 4, -st 60, -st 10 all with the same result; in fact only with -hpcoils 4 & -st 4 can I get more information (as above). I suspect the HPI coils failed at the indicated time points and the head location cannot be determined for an appropriate duration in the st buffer. Is this correct? More importantly, is there a fix? apart from not doing movement compensation.

Thanks in advance
Kambiz




------------------------------------------------------------
Kambiz Tavabi PhD
Institute for Learning & Brain Sciences
1715 Columbia Road N
Portage Bay Building
Box 357988
University of Washington
Seattle, WA 98195-7988
Tel: 206-685-6173
------------------------------------------------------------

________________________________
Please consider the environment before printing this e-mail.

The contents of this e-mail message (including any attachments) are confidential to and are intended to be conveyed for the use of the recipient to whom it is addressed only. If you receive this transmission in error, please notify the sender of this immediately and delete the message from your system. Any distribution, reproduction or use of this message by someone other than recipient is not authorized and may be unlawful.