Hi Daniel,

if you "only" want to do realign & unwarp (without having acquired a 
fieldmap), then this will realign the images and additionally correct 
for motion*B0 interaction effects. It will not undo the distortions 
induced by using EPI in the first place.

If you want to do that, you need an additional set of images coming from 
a dedicated fieldmap sequence. If you have those images, you can/must 
enter the TE of the early and late echo of the fieldmap, not the EPI 
(click Toolbox, Fieldmap), which will then allow you to properly unwarp 
the EPI images (a kind of "realign & unwarp & unwarp"). I don't think 
field strength is explicitly used anywhere as the fieldmap must of 
course be from the same subject, session, and thus, scanner.

Hope this helps,
Marko

Daniel ROQUET wrote:
> Hello SPM'ers
> I have two EPI data-sets (with distinct TE) which I want to realign and
> unwarp. I wonder whether FieldMap-Toolbox takes (1) the TE of the EPI
> images which have to be corrected, and (2) the strength of the magnetic
> field, into account . If it doesn't, why ? if it does, where can we
> change these parameters ?
> Best wishes
> --
> Daniel ROQUET
> PhD student
> Clinique Psychiatrique
> Unité INSERM 666
> 1 Place de l'Hôpital
> 67091 Strasbourg

-- 
____________________________________________________
PD Dr. med. Marko Wilke
  Facharzt für Kinder- und Jugendmedizin
  Leiter, Experimentelle Pädiatrische Neurobildgebung
  Universitäts-Kinderklinik
  Abt. III (Neuropädiatrie)


Marko Wilke, MD, PhD
  Pediatrician
  Head, Experimental Pediatric Neuroimaging
  University Children's Hospital
  Dept. III (Pediatric Neurology)


Hoppe-Seyler-Str. 1
  D - 72076 Tübingen, Germany
  Tel. +49 7071 29-83416
  Fax  +49 7071 29-5473
  [log in to unmask]

  http://www.medizin.uni-tuebingen.de/kinder/epn
____________________________________________________