Another query to my
colleagues on the mailbase please regarding our IL GEM
4000's;-
Firstly, we are
still having problems with cartridge failures on our 4000
analysers. Ours seem to be a mixture of initial CVP failures or a
parameter "fixing" which never resolves (resulting in a crtridge having to be
replaced), or various solution A//B errors. We have been told that no
other sites have this many problems. I wondered if this were true,
particularly as factors such as CVP failures could only be due to us installing
the cartridge in correctly, which is actually difficult to do wrong. This
means there is a consdierable down time when these analysers need to have their
cartridge replaced, the data copied, and be re-calibrated.
Secondly, we have
recently had an issue on SCBU where some completely erroneous results were
obtained on 3 separate occasions, but when the internal data for the analyser
was reviewed, nothing was amiss, according to IL. I have wondered whether
this might be due to the thick film technology that I understand is used on
these sensors, and may be due to the presence of free haemoglobin?
This could have arisen as a result of haemolysis while taking
these (capillary) samples, but the if this was the case, I still am
extremely concerned that result such as undetectable glucose, lactate, and
inappropriate pO2 and pH levels could be reported and potentially acted
upon.
If any of you have
any experiences or could pass this e-mail on to colleagues who may be involved,
I would be most grateful.
Many
Thanks
Pauline
Pauline Angel
Principal
Biochemist
Chemical Pathology
Department
Barnet & Chase Farm
Hospitals NHS Trust
Barnet Site
Wellhouse
Lane
Barnet
Herts
EN5 3DJ
Tel: 020 8216
4338
Patients first and foremost
The information contained in this email and any attachments may be subject to public disclosure under the NHS Code
of Openness or the Freedom of Information Act 2000. Unless the information is legally exempt from disclosure, the
confidentiality of this email and your reply cannot be guaranteed. The information transmitted is intended only for
the person to whom it is addressed and may contain confidential and/or privileged material. Any review,
retransmission, dissemination or other use of, or taking of any action in reliance upon, this information by persons
other than the intended recipient is strictly prohibited and may be unlawful. If you received this in error, please
contact and advise the sender and delete all record of the material from your system. Any views or opinions presented
are solely those of the author and do not represent the views of Barnet and Chase Farm Hospitals NHS Trust unless
otherwise specifically stated. Although believed to be virus free, accurate and complete, responsibility for any
loss or cost arising from its receipt or use or its incomplete or inaccurate transmission is hereby excluded to the
fullest extent possible.
|
------ACB discussion List Information--------
This is an open discussion list for the academic and clinical
community working in clinical biochemistry.
Please note, archived messages are public and can be viewed
via the internet. Views expressed are those of the individual and
they are responsible for all message content.
ACB Web Site
http://www.acb.org.uk
List Archives
http://www.jiscmail.ac.uk/lists/ACB-CLIN-CHEM-GEN.html
List Instructions (How to leave etc.)
http://www.jiscmail.ac.uk/