Print

Print


Oops. Jumped the gun there. That's not the cost function value. Sorry. I'll have another look...

On 9 Jan 2008, at 15:05, Dave Flitney wrote:

Kent,

Looking at the sources: in find_cost_minima it will output "COST minima" if verbose level is >= 3. Haven't had a chance to test it but hopefully that'll do the trick?

On 9 Jan 2008, at 14:45, Kent Conover wrote:

Hi Everybody,

Thanks to some advice that I received from Heather Urry, I have had some success unwarping 
my EPI volumes. My question is, is there a way to statistically assess the effectiveness of FUGUE? 
I imagine that one could compare the mutual information scores for the registrations of 
distorted and corrected volumes to the subjects high resolution anatomical. However, FLIRT 
apparently does not report the final values of the cost function. I would appreciate any 
suggestions.

Regards,

-Kent

PS. Here is Heather's advice on how to the transform the field map produced by the Siemens Magnetom Trio gre_field_mapping protocol for use by FUGUE:

Hi Kent,

For what it's worth, when I reconstruct our Siemens Magnetom Trio DICOM 
fieldmap images using MRIcron, I get 2 magnitude images and one phase map. 
The phase map is scaled with slope of 2 and an intercept of -4096. To 
convert that to radians per sec, I use the following command:

fslmaths fieldmapping64s006a201.nii.gz -mul 3.14159 -div 4096 -div .00246 
fieldmap.nii.gz

fieldmapping64s006a201.nii.gz is the name of the phase map.

Multiplying the phase map by 3.14159 and dividing by 4096 converts the 
image to -pi to pi. Dividing again by .00246 converts to 
radians/sec.  .00246 is the TE difference, in seconds. In my case, TE1 = 
2.84 ms and TE2 = 5.3 ms, thus TE diff = 2.46 ms, or .00246 s. 

fieldmap.nii.gz is the properly scaled fieldmap that is output by this 
command. It's the one that I use for B0 unwarping in FEAT.

Hope this is helpful!  Ignore if this isn't what you were looking for. :)

Cheers,

Heather





--
Cheers, Dave

Dave Flitney, IT Manager
Oxford Centre for Functional MRI of the Brain
E:[log in to unmask] W:+44-1865-222713 F:+44-1865-222717



--
Cheers, Dave

Dave Flitney, IT Manager
Oxford Centre for Functional MRI of the Brain
E:[log in to unmask] W:+44-1865-222713 F:+44-1865-222717