Dear all
when sending data over from the Avanto to the Leonardo and then to
the Mac Mini, make *sure* that there is enough space on the Leonardo
(and mac mini). I had not cleaned out the Leonardo for a while, and
there was a lack of space, which put up a warning on the Syngo
(Siemens) environment. Disturbingly, all of the files *looked* like
they had copied over (e.g., they looked fine on the patient viewer)
but when time came for reconstruction using ima2brik, things didn't
work. After I got rid of some files on Leonardo in Temp, I
transferred again from Avanto to Leonardo, and then all was fine.
So in short, make sure there is enough space for your data (we will
try to keep things cleaned up), and beware data corruption ides of
March.