Dear Elisabetta,
The trigger signal of the scanner is way too short to be detected by you
hardware. This is due to the limited sampling frequency of you hardware. You
need to extend the trigger signal so that it can be sampled adequately by
you hardware (serial port/parallel port/something else). I've done this
using a flipflop in feedback. Every time a trigger pulse is generated, the
flipflop changes its output state. Several flipflops in cascade can act as a
signal divider.
It requires some electronics skills. If you want to, I will make a easy
circuit scheme and post it, but it still needs to be done by a technician.
Greetings
Pieter Vandemaele